Lab Moves, Closures, or Transfers
Principal Investigators must submit an Amendments to Biosafety Protocols to the Institutional Biosafety Committee to:
- Document new laboratory locations
- Close-out old locations
- Transfer biological materials to a new PI.
All equipment must be decontaminated prior to removal from the laboratory using a
disinfectant appropriate for the agents used in the laboratory.
Biosafety Protocol
Biosafety Cabinets (BSCs)
- BSCs may need to be decontaminated using formaldehyde gas decontamination of HEPA
filters and/or wrapped prior to relocating or going to surplus. Call the Biological
Safety Office to determine the method of decontamination needed and to arrange for
this service.
- Any BSCs to be relocated must be identified and the new proposed location information
provided to the Biological Safety Office. We will perform a site assessment of the
new location to ensure the location complies with the CDC and NIH standards.
- Once moved, all BSCs must be recertified before re-use. Contact us to schedule recertification.
Biological Materials
All biologicals should be well-contained in sealed, leak-proof, durable, closed containers
prior to removal from the laboratory. If shipping biological materials to a new location,
the shipper will need to have completed shipping training and the shipments will need to be packaged appropriately.
Remember, biological materials must be either:
- Destroyed.
- Disposed of properly as Biological Waste (To request biowaste containers call 706-721-2434).
- Transferred from one laboratory to another laboratory, to a new off-campus location, or to a new
PI.
To maintain the integrity of any biological materials that will remain in storage
(i.e., in freezers) during relocation:
- Remove all breakable items from equipment.
- Secure tubes and loose objects in lidded boxes/containers to prevent loss, opening
or spreading of items.
- Pack empty spaces in freezers with foam rubber, cardboard or newspaper to prevent
shifting during a move. Some equipment may have to be placed on its side.
- Close and secure the doors of all equipment to prevent inadvertent opening during
moves (please lock or use packing tape).
- Decontaminate the outside of the equipment using a disinfectant appropriate for the
agents used in the laboratory.
Incubators
Drain and decontaminate incubators, if they are to be moved.
Additional equipment
All remaining equipment and surfaces in the old laboratory areas must also be decontaminated
using a disinfectant appropriate for the agents used in the laboratory.
Completing your move
- Notify the Biological Safety Office when all above procedures have been completed
so that we can fill out a Laboratory Clearance/ Transfer Form for the old laboratory. This form will need to be completed jointly with the Chemical
Safety Office.
- Once your laboratory is set up (Your equipment is in place, biohazard waste containers
are ready to go, BSCs certified), Please make arrangements with the Biological Safety
Office for a Laboratory Assistance Visit.
Download Lab moves and closeouts document